User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: KEITHLEY 160B DMM DATE: 15-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 22 NUMBER OF LINES: 48 CONFIGURATION: Fluke 5100B ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Check for any obvious physical damage. 1.001 DISP Check that all moving parts operate firmly and smoothly. 1.001 DISP Check for proper fuse size. 1.001 DISP Depress power switch and allow the T.I. 5 min.warm-up. 1.002 HEAD DC VOLTAGE TESTS 1.003 DISP select the 1mV range and short the input terminals. 1.003 DISP Adjust the front panel Zero Control for .000 display. 1.004 5100 1 1mV 0.2% 2W 2.001 5100 10 10mV 0.2% 2W 3.001 5100 100 100mV 0.2% 2W 4.001 5100 1 1V 0.2% 2W 5.001 5100 10 10V 0.2% 2W 6.001 5100 100 100V 0.2% 2W 7.001 5100 1000 500V 0.2% 2W 8.001 HEAD RESISTANCE TESTS 8.002 DISP Select the 1 ohm range and short the inputs. 8.002 DISP Adjust the front panel ZERO for .000 display 8.003 RESF 1 1Z 0.5% 2W 9.001 RESF 10 10Z 0.4% 2W 10.001 RESF 100 100Z 0.4% 2W 11.001 RESF 1 1KZ 0.2% 2W 12.001 RESF 10 10KZ 0.4% 2W 13.001 RESF 100 100KZ 0.4% 2W 14.001 RESF 1 1MZ 0.4% 2W 15.001 RESF 10 10MZ 0.4% 2W 16.001 HEAD DC CURRENT 16.002 5100 10 10uA 0.3% 2W 17.001 5100 100 100uA 0.3% 2W 18.001 5100 1 1mA 0.3% 2W 19.001 5100 10 10mA 0.3% 2W 20.001 5100 100 100mA 0.4% 2W 21.001 5100 1000 1000mA 0.4% 2W 22.001 DISP Performance verification complete. 22.001 DISP Perform any and all preventive maintenance and check 22.001 DISP off appropriate boxes on the TECCAM card. 22.002 END